The purpose of this investigation was to assess the effect of heparin coating of a new stent construction (Stent Graft, Jomed Implantate GmbH, Germany) on platelet and coagulation activity.
Methods: Stent grafts with an ePTFE membrane interfoliated between two stents were deployed in tubings to form Chandler loops. Fresh human blood with a low concentration of heparin was rotated for 1 h, then collected and used for measurements of platelet number, thrombin-antithrombin complex (TAT), CD11b, C3a and C5b-9. There were five study groups: Group 1, conventional unmodified stents (n = 8); Group 2, untreated stent grafts (n = 8); Group 3, heparin-coated stents and untreated membrane (n = 7); Group 4, heparin-coated stents and membrane (n = 8); Group 5, heparin-coated PVC tubings with no stents (n = 8).
Results: There was a significant drop in platelet count, increase in TAT-values and CD11b expression in Groups 1-3 but not in Group 4 compared to Group 5. Examination by scanning electron microscopy revealed extensive activation on non-modified stents but almost no deposition of thrombotic material on heparin-modified stent grafts.
Conclusions: With unmodified stents and membrane there were signs of significant activation of platelets and coagulation. In contrast, the heparin-coated stent graft induced much less alterations, indicating improved blood compatibility.