Recalibration of the point-of-care test for CARDIAC T Quantitative with Elecsys Troponin T 3rd generation

Clin Chim Acta. 2001 May;307(1-2):197-203. doi: 10.1016/s0009-8981(01)00428-4.

Abstract

The rapid troponin T assay CARDIAC T Quantitative was recalibrated using Elecsys Troponin T 3rd Generation as a new reference method. This paper presents the method comparisons at six centres using the new reference method. Method comparison between CARDIAC T Quantitative versus Elecsys Troponin T 3rd Generation were performed using 319 samples from patients with acute coronary syndromes. The quality of the CARDIAC T Quantitative was controlled by a daily single determination of CARDIAC Control Troponin T, and for the Elecsys Troponin T 3rd Generation, the Elecsys controls were included in each run. The results for the control materials for the CARDIAC T Quantitative were between 93% and 107% of the target values. The CV ranged from 7% to 16%. From the regression analysis, according to Bablok and Passing (y=1.07x) and the Bland and Altman plot, the bias between CARDIAC T Quantitative and Elecsys Troponin T 3rd Generation is from +6% to +7%. The correlation coefficient is 0.93, and a 3x3 comparison of the clinical efficiency yielded 92% clinical concordance between CARDIAC T Quantitative and Elecsys Troponin T 3rd Generation. In conclusion, CARDIAC T Quantitative was in good agreement with the reference and calibration method Elecsys Troponin T 3rd Generation.

MeSH terms

  • Calibration
  • Clinical Chemistry Tests / standards*
  • Humans
  • Point-of-Care Systems / standards*
  • Quality Control
  • Reference Standards
  • Troponin T / blood*
  • Troponin T / standards