Background: High-dose busulfan is widely used in conditioning regimens before hematopoietic stem cell transplantation in both adults and children. Large interindividual variability in pharmacokinetics after oral administration has been reported; therefore, therapeutic drug monitoring of busulfan may decrease the incidence of drug-related toxicity (for example, hepatic venoocclusive disease) and may also improve therapeutic efficacy.
Methods: Busulfan concentrations were quantified using 200 microL of plasma and liquid-liquid extraction with diethyl ether after the addition of [2H8]busulfan as the internal standard. Separation and detection of busulfan and [2H8]busulfan were achieved with a LUNA C8 column (5 microm; 150 x 2 mm i.d.) at 30 degrees C, a HP 1100 liquid chromatography system, and a HP 1100 single-quadrupole mass spectrometer. Busulfan and [2H8]busulfan were detected as ammonium adducts in selected-ion monitoring mode at m/z 264.2 and 272.2, respectively.
Results: The calibration curve was linear at 5-2000 microg/L busulfan. Intra- and interassay imprecision (CV) and bias were both <11%. The limits of detection and quantification were 2 and 5 microg/L, respectively. Extraction recovery of busulfan was >87%. Analysis of pharmacokinetics in four patients receiving high-dose busulfan indicated that minimum busulfan concentrations before the next dose were 405-603 microg/L, with no interference observed.
Conclusions: The new rapid and sensitive liquid chromatographic-mass spectrometric assay is an appropriate method for quantification of busulfan in human plasma, making therapeutic drug monitoring of busulfan faster and easier in clinical practice.