The performance of three vacuum-mixing cement guns--a comparison of the fatigue properties of simplex P cement

Int Orthop. 2001;25(5):290-3. doi: 10.1007/s002640100266.

Abstract

We tested three commercially available cement-mixing systems (Howmedica, Summit and Stryker). After mixing the cement was allowed to cure in metal moulds under standard conditions. Sixty-nine cement blocks were produced. The fatigue behavior of the blocks was tested until failure. A wide scattering of results was observed in all the three systems. Uneven mixing of the polymer and monomer was observed in the Howmedica system. A significantly stronger cement (P=0.018) was produced with the Stryker system.

Nous avons testé trois systèmes commercialement disponibles pour préparation du ciment (Howmedica, Summit et Stryker). Après mélange, le ciment a été déposé dans des moules métalliques sous environnement standardisé. 69 blocs de ciment ont été produits. Le comportement à la fatigue des blocs a été testé jusqu'à la rupture. Une dispersion large a été observée dans les trois systèmes. Un mélange irrégulier de polymère et de monomère a été observé dans le système Howmedica. Un ciment notablement plus résistant (P=0.018) a été produit avec le système Stryker.