Ultrasonic spectral parameter characterization of apoptosis

Ultrasound Med Biol. 2002 May;28(5):589-97. doi: 10.1016/s0301-5629(02)00492-1.

Abstract

Ultrasound (US) spectral analysis methods are used to analyze the radiofrequency (RF) data collected from cell pellets exposed to chemotherapeutics that induce apoptosis and other chemicals that induce nuclear transformations. Calibrated backscatter spectra from regions-of-interest (ROI) were analyzed using linear regression techniques to calculate the spectral slope and midband fit. Two f/2 transducers, with operating frequencies of 30 and 34 MHz (relative bandwidths of 93% and 78%, respectively) were used with a custom-made imaging system that enabled the collection of the raw RF data. For apoptotic cells, the spectral slope increased from 0.37 dB/MHz before drug exposure to 0.57 dB/MHz 24 h after, corresponding to a change in effective scatterer radius from 8.7 to 3.2 microm. The midband fit increased in a time-dependent fashion, peaking at 13dB 24 h after exposure. The statistical deviation of the spectral parameters was in close agreement with theoretical predictions. The results provide a framework for using spectral parameter methods to monitor apoptosis in in vitro and in in vivo systems and are being used to guide the design of system and signal analysis parameters.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Apoptosis / physiology*
  • Cells / diagnostic imaging*
  • Cells / ultrastructure
  • Deoxyribonucleases / metabolism
  • Humans
  • Models, Theoretical
  • Spectrum Analysis
  • Ultrasonics
  • Ultrasonography

Substances

  • Deoxyribonucleases