Stent deployment within a guide catheter aids removal of a fractured buddy wire

Catheter Cardiovasc Interv. 2002 Jun;56(2):212-4. doi: 10.1002/ccd.10193.

Abstract

We report a case of fractured buddy wire that was successfully removed by deploying a stent within the guide catheter, trapping the fractured segment of wire between stent and endoluminal surface of the guide catheter. This technique provides an alternative to either percutaneous snare or surgical intervention.

Publication types

  • Case Reports

MeSH terms

  • Angioplasty, Balloon, Coronary / adverse effects*
  • Cardiac Catheterization / adverse effects*
  • Coronary Angiography
  • Coronary Stenosis / diagnostic imaging
  • Device Removal*
  • Humans
  • Male
  • Middle Aged
  • Radiography, Interventional
  • Stents*