Rapid detection of methicillin resistance in coagulase-negative Staphylococci with the VITEK 2 system

J Clin Microbiol. 2002 Sep;40(9):3291-5. doi: 10.1128/JCM.40.9.3291-3295.2002.

Abstract

The aim of the present study was to evaluate the accuracy of the new VITEK 2 system (bioMérieux, Marcy l' Etoile, France) for the detection of methicillin resistance in coagulase-negative staphylococci (CoNS) by using AST-P515 and AST-P523 test cards. Analyses of the VITEK 2 oxacillin MIC determination evaluated according to the actual breakpoint (>/=0.5 micro g/ml) of the National Committee for Clinical Laboratory Standards resulted in a high sensitivity of 99.2% but a moderate specificity of 80%. The newly included oxacillin resistance (OR) test of the VITEK 2 system displayed a high sensitivity and a high specificity of 97.5 and 98.7%, respectively. Concordance between the results of the mecA PCR and the VITEK 2 oxacillin MIC was observed for almost all Staphylococcus epidermidis strains, but the reduced specificity was attributable to higher oxacillin MICs for mecA-negative non-S. epidermidis strains, especially S. saprophyticus, S. lugdunensis, and S. cohnii. Evaluation of alternative oxacillin MIC breakpoints of 1, 2, or 4 micro g/ml resulted in improved degrees of specificity of 84, 90.7, and 97.3%, respectively. Only minor changes occurred in the corresponding sensitivity values, which were 98.4, 97.5, and 97.5%, respectively. Methicillin resistance in CoNS was detected after 7 and 8 h in 91.1 and 93.5% of the mecA-positive strains, respectively, by the VITEK 2 OR test and in 86.3 and 89.5% of the mecA-positive strains, respectively, by VITEK 2 oxacillin MIC determination. After 7 and 8 h the VITEK 2 OR test classified 59.2 and 78.9% of the mecA-negative strains, respectively, as susceptible to oxacillin, whereas comparable values were obtained 2 h later by VITEK 2 oxacillin MIC determination. The results of our study encourage the use of the VITEK 2 system, which proved to be a highly reliable and rapid phenotypic method for the detection of methicillin resistance in CoNS.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Bacterial Proteins*
  • Carrier Proteins
  • Coagulase / metabolism*
  • Hexosyltransferases*
  • Humans
  • Methicillin Resistance*
  • Microbial Sensitivity Tests / methods
  • Microbial Sensitivity Tests / standards
  • Muramoylpentapeptide Carboxypeptidase
  • Oxacillin / pharmacology
  • Penicillin-Binding Proteins
  • Penicillins / pharmacology
  • Peptidyl Transferases*
  • Reagent Kits, Diagnostic*
  • Sensitivity and Specificity
  • Staphylococcal Infections
  • Staphylococcus / drug effects*
  • Time Factors

Substances

  • Bacterial Proteins
  • Carrier Proteins
  • Coagulase
  • Penicillin-Binding Proteins
  • Penicillins
  • Reagent Kits, Diagnostic
  • Peptidyl Transferases
  • Hexosyltransferases
  • Muramoylpentapeptide Carboxypeptidase
  • Oxacillin