Multiple gradient echo sequence optimized for rapid, single-scan mapping of R(2)(*) at high B0

Magn Reson Med. 2002 Nov;48(5):867-76. doi: 10.1002/mrm.10291.

Abstract

A multiple-gradient-echo sequence is proposed for accurately mapping R(2)(*) in the presence of in-slice macroscopic susceptibility gradients. In-slice signal loss caused by background macroscopic susceptibility gradients is mitigated by combining three successive gradient-echo images whose slice refocus gradients are successively incremented. The optimum incrementation of slice-refocusing gradients was determined by numerical simulation. By repeating further cycles of three images in the same sequence, artifact-compensated data spanning a range of echo times (TEs) was acquired leading to single-scan, R(2) (*) maps that are quantitatively reflective of microscopic field inhomogeneities. The performance of the sequence was demonstrated at 3.0T, first with a doped aqueous phantom, and then on the head of a normal volunteer. That performance is compared quantitatively with previously published work.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Artifacts
  • Brain / anatomy & histology*
  • Humans
  • Magnetic Resonance Imaging / methods*