Development of a solid-phase cleanup and portable rapid flow-through enzyme immunoassay for the detection of ochratoxin a in roasted coffee

J Agric Food Chem. 2002 Nov 20;50(24):6964-7. doi: 10.1021/jf020553o.

Abstract

A membrane-based flow-through enzyme immunoassay (patent application pending) for the detection of ochratoxin A (OA) in roasted coffee was developed. First, an extraction and solid-phase cleanup method was developed. A high partition coefficient for OA in the mobile phase was achieved by using methanol/5% aqueous NaHCO(3) as the sample extraction and cleanup solvent. The solid-phase (aminopropyl) cleanup was developed to chromatographically elute OA but retain cross-reacting compounds. Without using aminopropyl cleanup, cross-reacting compounds resulted in 100% false positives for both flow-through enzyme immunoassay and HPLC methods. However, after cleanup with aminopropyl, no false positives were observed. The flow-through results were visually evaluated. The sensitivity achieved for the flow-through was 4 microg kg(-1) in spiked roasted coffee. The assay was used to screen roasted coffee samples. Results were confirmed with HPLC with a detection limit of 1 microg kg(-1).

Publication types

  • Comparative Study

MeSH terms

  • Chromatography, High Pressure Liquid
  • Coffea / chemistry*
  • False Positive Reactions
  • Hot Temperature*
  • Immunoenzyme Techniques / methods*
  • Ochratoxins / analysis*
  • Seeds / chemistry*
  • Sodium Bicarbonate
  • Solvents

Substances

  • Ochratoxins
  • Solvents
  • ochratoxin A
  • Sodium Bicarbonate