Changes in nerve conduction velocity across the elbow due to experimental error

Muscle Nerve. 2002 Dec;26(6):838-40. doi: 10.1002/mus.10259.

Abstract

One diagnostic criterion for ulnar nerve mononeuropathy at the elbow (UNE) is a decrease in across-elbow nerve conduction velocity (NCV) > 10 m/s compared to the forearm segment. Distance and latency measurement errors are an inherent part of NCV calculations. Twenty electromyographers measured the latencies of stored ulnar compound muscle action potentials and measured the forearm and across-elbow distances along the ulnar nerve. Based on previously published equations, experimental error in NCV was calculated for various NCVs. The mean distances and standard deviations for the forearm and elbow segments were 212.5 +/- 2.1 mm and 86.7 +/- 4.2 mm, respectively. For an NCV of 55 m/s, a difference of 14 m/s between the two segments can occur from measurement error alone. Distance measurements about the elbow are fraught with interobserver errors rendering the resultant NCV of that segment of limited value as a sole criterion for the diagnosis of UNE.

MeSH terms

  • Artifacts
  • Diagnostic Errors*
  • Humans
  • Mononeuropathies / physiopathology*
  • Neural Conduction / physiology*
  • Research Design
  • Ulnar Nerve / physiopathology*
  • Ulnar Nerve Compression Syndromes / physiopathology*