Fourteen titanium dental implants (Tioblast) were implanted singly in the proximal tibia of New Zealand rabbits for 120 days. A bone defect was surgically produced and filled with Bio-Oss around six of these implants. After the animals were sacrificed and their organs harvested, bone segments were fixed and methacrylate embedded after the push-in test had been performed. Microradiography was performed on longitudinal sections of the implants, whereas scanning electron microscope analysis was performed on the remaining embedded half-implants using secondary electrons only. The results showed that the implants were apically and coronally surrounded by bone, whether Bio-Oss was used or not. Fractures were evident through the newly formed bone and between the pre-existing and newly formed bone. Some fracture lines propagated through the bone and stopped at the implant surface without continuing along the bone-titanium interface. Detachment between the implant and the bone occurred at the coronal extremity of the implants and along its cervical region. These results highlight the fact that the bone-titanium interface has a high resistance to loading. It exhibited greater resistance than the newly formed bone and seems to behave in a manner similar to the cement lines of osteons.
Copyright 2002 Wiley Periodicals, Inc. J Biomed Mater Res 64A: 365-371, 2003