Purpose: To evaluate the properties of the AcrySof(R) SA30AL (Alcon Laboratories, Inc.) single-piece foldable posterior chamber intraocular lens (IOL).
Setting: Center for Research on Ocular Therapeutics and Biodevices, Storm Eye Institute, Medical University of South Carolina, Charleston, South Carolina, USA.
Methods: Two nonimplanted clinical-quality AcrySof IOLs were examined by gross, light, and scanning electron microscopy (SEM). In addition, 2 eyes implanted with this IOL obtained post-mortem, the first such eyes accessioned in our laboratory and the first described to date, were examined using the Miyake-Apple posterior photographic technique and by histologic sections.
Results: Scanning electron microscopy of the SA30AL IOL showed excellent surface finish. The edge of the optic was square (truncated) and had a matte (velvet or ground-glass) appearance, a feature that may minimize edge glare and other visual phenomena. A well-fabricated square or truncated optic edge was demonstrated. Miyake-Apple analysis revealed that the SA30AL IOL showed appropriate fit and configuration within the capsular bag. Histologic correlation of the IOL's square edge and its relation to the capsular bag and adjacent Soemmering's ring were noted.
Conclusions: The AcrySof SA30AL IOL is a well-fabricated lens that situates well in the capsular bag. The truncated optic and its relationship to adjacent structures show a morphological profile that has been shown to be highly efficacious in reducing the rate of posterior capsule opacification.