We performed a controlled laboratory study to evaluate the initial fixation strength of press-fit technique. Forty porcine lower limbs were used and divided into four groups according to the method of fixation; group 1 (press-fit+1.4 mm), in which the diameter difference between the bone plug and the femoral tunnel was 1.4 mm; group 2 (press-fit+1.4 mm, 30 degrees), in which the diameter difference was the same with group 1, but the tensile loading axis was 30 degrees away from the long axis of the femoral tunnel; group 3 (titanium), in which a titanium interference screw was used for fixation; group 4 (bioabsorbable), in which a bioabsorbable interference screw was used for fixation. The graft in the press-fit group was harvested with a hollow oscillating saw with inner diameter of 9.4 mm to obtain consistent and completely circular shape of the bone plug. The femoral tunnel with diameter of 8 mm was drilled at the original ACL insertion. Following the bone plug insertion into the femoral tunnel and applying a preload of 20 N, the specimen underwent 500 loading cycles between 0 and 2 mm of displacement. Thereafter the specimen was loaded to failure. There was no fixation site failure during the cyclic loading test. Significant differences in the stiffness, linear load, or failure mode among the groups were not found. The average ultimate failure load of group 1 and group 2 were not significantly different from those of group 3 and group 4. The press-fit groups demonstrated sufficient fixation strength for the rehabilitation and interference screw groups. The completely circular shape of the bone plug and increased diameter difference between the bone plug and the femoral tunnel seemed to contribute to the strong fixation.