Introduction: Catheter ablation has recently been used for curative treatment of atrial fibrillation.
Methods and results: Three of 239 patients who underwent ablation close to the pulmonary vein (PV) ostia at our institute developed severe hemoptysis, dyspnea, and pneumonia as early as 1 week and as late as 6 months after the ablation. Because the patients were arrhythmia-free, the treating physician initially attributed the symptoms to new-onset pulmonary disease (e.g., bronchopulmonary neoplasm). After absent PV flow was confirmed by transesophageal echocardiography, transseptal contrast injection depicted a totally occluded PV in all three patients. Successful recanalization, even in chronically occluded Pvs, was performed in all patients. During follow-up, Doppler flow measurements by transesophageal echocardiography demonstrated restenosis in all primarily dilated PV, which led to stent implantation.
Conclusion: PV stenosis/occlusion after catheter ablation of atrial fibrillation occurs in a subset of patients. However, because in-stent restenosis occurred in two patients after 6 to 10 weeks, final interventional strategy for PV stenosis or occlusion remains unclear. To prevent future PV stenosis or occlusion, a decrease in target temperature and energy of radiofrequency current or the use of new energy sources (ultrasound, cryothermia, microwave) seems necessary.