Increased diffusion sensitivity with hyperechos

Magn Reson Med. 2003 Jun;49(6):1098-105. doi: 10.1002/mrm.10457.

Abstract

It is shown that the introduction of a 180 degrees refocusing pulse into a standard diffusion weighted stimulated echo sequence is equivalent to the simplest hyperecho sequence with identical diffusion weighting but equal or greater signal-to-noise (SNR) and thus equal or greater diffusion contrast. For high b-value imaging, the hyperecho sequence thus possesses the high diffusion contrast in the presence of small T(1)/T(2) ratios characteristic of stimulated echo sequences but with less than the 50% loss in SNR that is associated with the stimulated echo. For low b-value imaging, the hyperecho signal converges to that of the standard spin echo. The advantages of the two-pulse diffusion weighted hyperecho sequence are demonstrated theoretically. Experimental results are shown in the application to high angular resolution diffusion encoding (HARD) in normal human brain.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, Non-P.H.S.
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Anisotropy
  • Brain / physiology*
  • Diffusion Magnetic Resonance Imaging*
  • Humans
  • Models, Theoretical