Keyhole Dixon method for faster, perceptually equivalent fat suppression

J Magn Reson Imaging. 2003 Jul;18(1):103-12. doi: 10.1002/jmri.10323.

Abstract

Purpose: To reduce the acquisition time associated with the two-point Dixon fat suppression technique by combining a keyhole in-phase (Water + Fat) k-space data set with a full out-of-phase (Water - Fat) k-space data set and optimizing the keyhole size with a perceptual difference model.

Materials and methods: A set of keyhole Dixon images was created by varying the number of lines in the keyhole data set. Off-resonance correction was incorporated into the image reconstruction process to improve the homogeneity of the fat suppression. A perceptual difference model (PDM) was validated with human observer experiments and used to compare the keyhole images to images from a full two-point Dixon acquisition. The PDM was used to determine the smallest keyhole width required to obtain perceptual equivalence to images obtained from the full two-point Dixon method.

Results: In experimental phantom studies, the keyhole Dixon image reconstructed from 96 of 192 Water + Fat k-space lines and 192 Water - Fat k-space lines was perceptually equivalent to the full (192 + 192) two-point Dixon images, resulting in a 25% reduction in scan time. Clinical images of a volunteer's knee, orbits, and abdomen created from the smallest, perceptually equivalent keyhole width resulted in a 27%-38% reduction in total scan time.

Conclusion: This method improves the temporal efficiency of the conventional two-point Dixon technique and may prove especially useful for high-field systems where specific absorption rate (SAR) limits will constrain radiofrequency (RF)-based fat suppression techniques.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Adipose Tissue
  • Humans
  • Image Processing, Computer-Assisted / methods*
  • Magnetic Resonance Imaging / methods*
  • Phantoms, Imaging