Feasibility of an implantable arrhythmia monitor

Pacing Clin Electrophysiol. 1992 Dec;15(12):2232-5. doi: 10.1111/j.1540-8159.1992.tb04164.x.

Abstract

Conventional Holter monitoring is of limited benefit in patients with infrequent symptoms suspected to be related to arrhythmia. A small recorder implanted subcutaneously might obviate many limitations of conventional monitoring. To determine the feasibility of obtaining adequate electrocardiographic signals from such a device, a prototype was temporarily implanted in 17 patients undergoing pacemaker implantation. The prototype contained four disc-shaped titanium electrodes, 0.21 inches in diameter embedded in epoxy. The four electrodes were in a square configuration spaced 0.72 inches center to center and were placed face down in a subcutaneous pocket in the left pectoral region. Bipolar recordings were made from a horizontal pair, a vertical pair, and both diagonal paris of electrodes (interelectrode distance 1.02 inches) and recorded on electromagnetic tape after filtering at 0.5-250 Hz. The mean peak-to-peak amplitude in each configuration was determined over a five-beat interval. Clear recordings were obtained from all 17 patients with recognizable P, QRS, and T waves. The amplitude of the signals obtained from the diagonal pairs of electrodes (175 +/- 51 and 170 +/- 54 microV) were greater than obtained from either the vertical pair (142 +/- 62 microV, P = 0.08 compared to diagonal electrodes) or the horizontal pair of electrodes (105 +/- 54 microV, P < 0.01). The maximum amplitude recorded from any configuration was 189 +/- 54 microV. In six patients the device was also tested with the electrodes face up in the subcutaneous pocket.(ABSTRACT TRUNCATED AT 250 WORDS)

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Arrhythmias, Cardiac / complications
  • Arrhythmias, Cardiac / diagnosis*
  • Electrocardiography, Ambulatory / instrumentation*
  • Electrodes, Implanted
  • Equipment Design
  • Feasibility Studies
  • Female
  • Humans
  • Male
  • Prostheses and Implants*
  • Syncope / etiology