An accelerometer-based rate adaptive generator (EXCEL VR) has been introduced. A preclinical group of 22 subjects with strap-on devices was observed and reported. A clinical protocol including observation of rate adaptive response to typical daily activities and incremental exercise on a treadmill was administered in seven implanted patients. Indications for implantation in these patients was either second- or third-degree atrioventricular block (five patients, VVIR pacing mode) and sick sinus syndrome (two patients, AAIR pacing mode). Mean pacing rates were 50 ppm (supine), 56 ppm (standing), 77 ppm (descending the stairs), 81 ppm (slow walk), 83 ppm (slow stair climb), 91 ppm (fast walk), and 92 ppm (fast stair climb). When the arm proximal to the pulse generator was exercised, the rate rose to 92 ppm. When the distal arm was strained, the rate was 63 ppm. During treadmill testing, rates between 82 ppm (2 km/hour) and 104 ppm (5 km/hour) were observed. This accelerometer-based rate adaptive pulse generator provided a proportional response to graded activities of treadmill exercise and daily living in these groups of preclinical and clinical subjects.