Testosterone measured by 10 immunoassays and by isotope-dilution gas chromatography-mass spectrometry in sera from 116 men, women, and children

Clin Chem. 2003 Aug;49(8):1381-95. doi: 10.1373/49.8.1381.

Abstract

Background: Commercially available testosterone immunoassays give divergent results, especially at the low concentrations seen in women. We compared immunoassays and a nonimmunochemical method that could quantify low testosterone concentrations.

Methods: We measured serum testosterone in 50 men, 55 women, and 11 children with use of eight nonisotopic immunoassays, two isotopic immunoassays, and isotope-dilution gas chromatography-mass spectrometry (ID/GC-MS).

Results: Compared with ID/GC-MS, 7 of the 10 immunoassays tested overestimated testosterone concentrations in samples from women; mean immunoassay results were 46% above those obtained by ID/GC-MS. The immunoassays underestimated testosterone concentrations in samples from men, giving mean results 12% below those obtained by ID/GC-MS. In women, at concentrations of 0.6-7.2 nmol/L, 3 of the 10 immunoassays gave positive mean differences >2.0 nmol/L (range, -0.7 to 3.3 nmol/L) compared with ID/GC-MS; in men at concentrations of 8.2-58 nmol/L, 3 of the 10 immunoassays tested gave mean differences >4.0 nmol/L (range, -4.8 to 2.6 nmol/L).

Conclusion: None of the immunoassays tested was sufficiently reliable for the investigation of sera from children and women, in whom very low (0.17 nmol/L) and low (<1.7 nmol/L) testosterone concentrations are expected.