Simultaneous determination of JTT-501 and its main metabolite in human plasma by liquid chromatography-ionspray mass spectrometry

J Chromatogr B Analyt Technol Biomed Life Sci. 2003 Aug 25;794(1):167-74. doi: 10.1016/s1570-0232(03)00453-7.

Abstract

An LC-MS-MS analytical method was developed for the determination of a new antidiabetic agent, JTT-501 and its main metabolite (JTP-20604) in human plasma. The compounds were isolated from plasma by protein precipitation before analysis by HPLC with atmospheric pressure positive ionisation MS-MS detection. An isotopically labelled analog of JTT-501 was used as the internal standard. Linearity was demonstrated over the calibration range of about 5-10000 ng/ml for both compounds. The assay was validated with respect to accuracy, precision and analyte stability. This method was used for the determination of plasma concentrations for the two compounds in a clinical tolerability study. A cross-validation exercise between two different mass spectrometers, used for the determination of clinical samples, is also reported.

MeSH terms

  • Chromatography, High Pressure Liquid / methods*
  • Humans
  • Hypoglycemic Agents / blood*
  • Isoxazoles / blood*
  • Reproducibility of Results
  • Spectrometry, Mass, Electrospray Ionization / methods*

Substances

  • Hypoglycemic Agents
  • Isoxazoles
  • JTT 501