To study the response of cultured cells to the cyclical biaxial biomechanical strain, a system was made according to the principle of Winston. The system was made of culture dish, pressure cavity and step motor. The culture dishes were made of polymethyl methacrylate and the bottom of the culture dishes were made of silicone rubber membrane. When the diameter of the silicone rubber tube and the distance between rollers and curvilinear plate changed, the pressure in the pressure cavity changed and the strain of the membrane changed from 0 to 4%. When the roll of the step motor changed, the frequency of the strain changed from 0.1 to 5 Hz.