[Ionic dialysance to control the dose of dialysis. One year experience]

Nefrologia. 2003 Sep-Oct;23(5):444-50.
[Article in Spanish]

Abstract

The Diascan equipment (Hospal) measures ionic dialysane which it derives the K and the Kt. If we divide the Kt obtained with Diascan between the Kt/V obtained by a simplified formula, it result a value of V for every patient. Entering this V in the Diascan software we can obtain a Kt/V (Diascan Kt/V), similar in theory to the simplified Kt/V. In the year 2002 we have controlled the delivered dialysis in our unit with the Diascan Kt/V. The aim of the present study was to study the agreement between de Diascan Kt/V and the Lowrie Kt/V. During the year 2002, 63 patients have been dialyzed in monitors with Diascan equipment. We calculated the V of each patient by dividing the Kt Diascan between the Lowrie Kt/V in the same dialysis session. The mea of the two consecutive measurements was considered the V value. Throughout the year 2002, 7 agreement studies were realized. The inter-method variability was assessed by the relative difference (absolute difference Diascan Kt/V-Lowrie Kt/V, divided by the average of both tests). A good agreement was considered when the relative difference was equal or lower than 10%. In the 7 agreement studies realized, the mean of the relative difference oscilled between 5.2 and 6.6%, and the percentage of patients with a relative difference equal or lower than 10% oscilled between 83 and 91%. During a month, the Diascan Kt/V was controlled in all dialysis sessions in 41 patients (554 sessions in total). Failure in the lecture of Kt/V Diascan was observed in 41 sessions (7%). A Diascan Kt/V greater than 1 (the minimum delivered dialysis considered in our unit) was obtained in 93% of the valid sessions. 38 of 41 patients had a mean monthly Diascan Kt/V greater than 1. The coefficient of variability of any patient oscilled between 2.1 and 12.4% (mean 5.1%). Diascan Kt/V is good procedure for the monitoring the delivered dialysis without blood sampling or any additional costs.

Publication types

  • English Abstract

MeSH terms

  • Evaluation Studies as Topic
  • Hemodialysis Solutions / chemistry*
  • Hemodialysis Solutions / metabolism
  • Humans
  • Ions
  • Kidney Failure, Chronic / therapy
  • Monitoring, Physiologic / methods*
  • Renal Dialysis / methods*
  • Urea / metabolism*

Substances

  • Hemodialysis Solutions
  • Ions
  • Urea