Multifiber, multiwavelength, fiber optic fluorescence spectrophotometer

IEEE Trans Biomed Eng. 1992 Sep;39(9):886-92. doi: 10.1109/10.256421.

Abstract

A fiberoptic spectrofluorimeter was constructed capable of remote operation of up to 18 biosensors under fully automated, multiplexed control. Excitation and emission wavelengths were separately adjustable, as were optical bandwidths, signal gain, phase, integration time, and illumination duty cycle. Synchronous demodulation was used to enhance signal-to-noise ratio and allow operation under ambient lighting conditions. Detection limit for aqueous sodium-fluorescein was below 1 ng/mL at greater than unity signal-to-noise ratio, with effective sampling volumes less than 0.02 microL. The fluorimeter showed characteristics fiberoptic sampling behavior with asymptotic nonzero signal magnitudes under increasing pathlength conditions.

MeSH terms

  • Equipment Design / standards
  • Evaluation Studies as Topic
  • Fiber Optic Technology / instrumentation
  • Fiber Optic Technology / methods
  • Fiber Optic Technology / standards*
  • Optical Fibers
  • Sensitivity and Specificity
  • Signal Processing, Computer-Assisted / instrumentation
  • Spectrometry, Fluorescence / instrumentation
  • Spectrometry, Fluorescence / methods
  • Spectrometry, Fluorescence / standards*