Rapid gas chromatography-mass spectrometry screening method for human pharmaceuticals, hormones, antioxidants and plasticizers in water

J Chromatogr A. 2004 Mar 12;1029(1-2):223-37. doi: 10.1016/j.chroma.2003.11.098.

Abstract

A rapid gas chromatography-mass spectrometry (GC-MS) method was developed and validated allowing quantification at the ng/l level of 19 analytes in water including human pharmaceuticals, hormones, antioxidants and a plasticizer. On-line continuous liquid-liquid extraction with dichloromethane of 10-401 unfiltered water samples was used to achieve a 10000-40000-fold concentration factor. No sample cleanup or derivatization was required. Recoveries ranged from 57 to 120%. Application of the method to water recycling plant effluent demonstrated the presence of nearly all targeted compounds at ng/l to microgram/l levels. Screening for nontarget compounds in the treated effluent samples indicated the method could be readily extended to include additional analytes.