A novel interface between high performance liquid chromatography (HPLC) and inductively coupled plasma-mass spectrometry (ICP-MS) is described. The eluent from the HPLC is nebulised into a heated cyclone spray-chamber and the solvent removed using a Nafion membrane drier, held at 75 degrees C, and a cryogenic condenser. The condenser consists of 6 Peltier heat pumps connected to liquid cooled aluminium blocks. At a nebuliser gas flow rate of 0.6 l min(-1), the membrane drier removes 58% of the vapour and the Peltier condenser 75% of the remaining vapour, i.e. a total desolvation efficiency of 89%. This enables the use of HPLC solvents which otherwise would destabilise the ICP, e.g. 100% acetonitrile or methanol, and permits the use of solvent gradients with minimal baseline drift. The system has been applied to the determination of platinum species in an organoplatinum drug used for chemotherapy in human plasma ultrafiltrate of patients treated with this new drug (JM-216). The limit of detection for platinum species has been 0.6 ng nl(-1) (i.e. 120 pg of Pt) and several species have been separated with good resolution.