Abstract
We have conducted an international quality assurance study of filovirus, Lassa virus, and orthopox virus PCR with 24 participants. Of the participating laboratories, 45.8 and 66.7% detected virus in all plasma samples, which contained > or = 5,000 and > or = 100,000 copies per ml, respectively. Sensitivity levels were not significantly different between viruses. False-negative results were attributable to a lack of sensitivity.
Publication types
-