Intelligent walkers for the elderly: performance and safety testing of VA-PAMAID robotic walker

J Rehabil Res Dev. 2003 Sep-Oct;40(5):423-31. doi: 10.1682/jrrd.2003.09.0423.

Abstract

A walker that could help navigate and avoid collisions with obstacles could help reduce health costs and increase the quality of care and independence of thousands of people. This study evaluated the safety and performance of the Veterans Affairs Personal Adaptive Mobility Aid (VA-PAMAID). We performed engineering tests on the VA-PAMAID to determine safety factors, including stability, energy consumption, fatigue life, and sensor and control malfunctions. The VA-PAMAID traveled 10.9 km on a full charge and avoided obstacles while traveling at a speed of up to 1.2 m/s. No failures occurred during static stability, climatic, or fatigue testing. Some problems were encountered during obstacle climbing and sensor and control testing. The VA-PAMAID has good range, has adequate reaction time, and is structurally sound. Clinical trials are planned to compare the device to other low-technical adaptive mobility devices.

Publication types

  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Equipment Design
  • Equipment Safety
  • Humans
  • Robotics
  • Walkers*