Initial data from the 30-element ORTEC HPGe detector array and the XSPRESS pulse-processing electronics at the SRS, Daresbury Laboratory

J Synchrotron Radiat. 1998 May 1;5(Pt 3):845-7. doi: 10.1107/S0909049597013125. Epub 1998 May 1.

Abstract

Following the completion of the collaborative project between CLRC Daresbury Laboratory and EG&G ORTEC to develop the world's first 30-element HPGe detector for fluorescence XAFS, it has now been tested and commissioned at the SRS. The system was commissioned with the XSPRESS digital pulse-processing electronics and this has demonstrated processed count rates in excess of 10 MHz. Initial data have been recorded and are presented.