Comparison of matched-filtered two-dimensional projection and elliptical centric-ordered three-dimensional contrast-enhanced magnetic resonance angiography

J Magn Reson Imaging. 2004 Sep;20(3):435-42. doi: 10.1002/jmri.20139.

Abstract

Purpose: To compare the image quality of matched-filtered two-dimensional projection magnetic resonance angiography (MRA) and elliptical centric-ordered (EC) three-dimensional MRA.

Materials and methods: Signal-to-noise ratios (SNRs) of matched-filtered two-dimensional projection and EC three-dimensional MRA are developed theoretically and compared by clinical studies, in which 10-20 mL of gadolinium (Gd) was injected at 1.5 mL/second. The artery-vein contrast in two-dimensional projection MRA was managed by manually selecting specific templates for the matched filters.

Results: The SNR of matched-filtered two-dimensional projection MRA is superior to that of EC three-dimensional MRA for vessels wider than one pixel due to the integral effect. The artery-vein contrast can be managed flexibly in two-dimensional projection MRA by choosing different templates for the matched filter, while the artery-vein contrast in EC three-dimensional MRA is solely determined by the timing to start the acquisition.

Conclusion: Matched-filtered two-dimensional projection MRA provides comparable image quality and is a flexible alternative to EC three-dimensional MRA in applications where contrast timing is difficult and temporal information is of interest.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Contrast Media
  • Gadolinium DTPA
  • Humans
  • Imaging, Three-Dimensional
  • Magnetic Resonance Angiography / methods*

Substances

  • Contrast Media
  • gadodiamide