Abstract

Online prostate positioning using gold markers and a standard video-based electronic portal imaging device is reported. The average systematic (random) errors have been reduced from 2.1 mm (2.7 mm) to 0.5 mm (1.5 mm) in AP direction, 1.1 mm (1.7 mm) to 0.7 mm (1.2 mm) SI and 1.2 mm (1.7 mm) to 0.6 mm (1.3 mm) LR.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms*
  • Electronics, Medical
  • Humans
  • Male
  • Prostatic Neoplasms / radiotherapy*
  • Radiotherapy / instrumentation
  • Silicon

Substances

  • Silicon