Persistent Acinetobacter baumannii? Look inside your medical equipment

Infect Control Hosp Epidemiol. 2004 Nov;25(11):1002-4. doi: 10.1086/502335.

Abstract

Two outbreaks of multidrug-resistant Acinetobacter baumannii occurred in our hospital. The outbreak strains were eventually isolated from respiratory ventilators, an apparatus used to cool or warm patients, and four continuous veno-venous hemofiltration machines. Removing dust from the machines and replacing all dust filters brought the outbreaks to an end.

MeSH terms

  • Acinetobacter Infections / drug therapy
  • Acinetobacter Infections / epidemiology*
  • Acinetobacter Infections / transmission*
  • Acinetobacter baumannii / isolation & purification*
  • Disease Outbreaks / statistics & numerical data*
  • Disease Transmission, Infectious / prevention & control
  • Disease Transmission, Infectious / statistics & numerical data
  • Drug Resistance, Multiple, Bacterial
  • Equipment Contamination / prevention & control
  • Equipment Contamination / statistics & numerical data*
  • Equipment and Supplies, Hospital / microbiology*
  • Humans
  • Infection Control / methods
  • Netherlands / epidemiology
  • Respiration, Artificial / instrumentation