Contrast-enhanced MR angiography at 1.5 T after implantation of platinum stents: in vitro and in vivo comparison with conventional stent designs

AJR Am J Roentgenol. 2005 Jan;184(1):288-94. doi: 10.2214/ajr.184.1.01840288.

Abstract

Objective: The objective of our study was to evaluate the in vitro and in vivo 3D contrast-enhanced MR angiography characteristics of a new platinum-based balloon-expandable stent system and compare this system with a variety of competing metallic stents.

Materials and methods: All experiments were performed on 1.5-T scanners. In vitro experiments were performed using 10 stents implanted into a custom-built phantom. Different orientations of the stents along the magnetic field and multiple flip angles were examined. In addition, 19 patients underwent contrast-enhanced MR angiography after the implantation of 36 stents, including four patients with six platinum stents. Angiographic correlation was available for all 19 patients, and luminal patency and stent-induced artifacts were assessed quantitatively.

Results: Of the tested balloon-expandable stents, only the platinum-based stents created artifact causing luminal narrowing of 30% or less. All other balloon-expandable stents induced larger artifacts that resulted in higher degrees of narrowing. Thus, if patent, the platinum-based stents allow significant in-stent stenosis to be ruled out reliably. Selected nitinol- or tantalum-based self-expandable stents also are suitable in this regard.

Conclusion: Of the tested devices, platinum-based stents are the only type of currently available balloon-expandable stent that creates 30% or less artifact-induced apparent stenosis and thus are suitable for MR angiography.

MeSH terms

  • Aged
  • Angiography
  • Arterial Occlusive Diseases / therapy*
  • Artifacts
  • Contrast Media
  • Female
  • Gadolinium DTPA
  • Humans
  • Image Processing, Computer-Assisted
  • Imaging, Three-Dimensional
  • In Vitro Techniques
  • Magnetic Resonance Angiography / methods*
  • Male
  • Middle Aged
  • Phantoms, Imaging
  • Platinum
  • Retrospective Studies
  • Stents*
  • Vascular Patency

Substances

  • Contrast Media
  • Platinum
  • gadodiamide
  • Gadolinium DTPA