Objectives: The aim of this study was to investigate the occurrence and the clinical impact of stent fractures after femoropopliteal stenting.
Background: The development of femoral stent fractures has recently been described; however, there are no data about the frequency and the clinical relevance.
Methods: A systematic X-ray screening for stent fractures was performed in 93 patients. In total, 121 legs treated by implantation of self-expanding nitinol stents were investigated after a mean follow-up time of 10.7 months. The mean length of the stented segment was 15.7 cm.
Results: Overall, stent fractures were detected in 45 of 121 treated legs (37.2%). In a stent-based analysis, 64 of 261 stents (24.5%) showed fractures, which were classified as minor (single strut fracture) in 31 cases (48.4%), moderate (fracture of >1 strut) in 17 cases (26.6%), and severe (complete separation of stent segments) in 16 cases (25.0%). Fracture rates were 13.2% for stented length < or =8 cm, 42.4% for stented length >8 to 16 cm, and 52.0% for stented length >16 cm. In 21 cases (32.8%) there was a restenosis of >50% diameter reduction at the site of stent fracture. In 22 cases (34.4%) with stent fracture there was a total stent reocclusion. According to Kaplan-Meier estimates, the primary patency rate at 12 months was significantly lower for patients with stent fractures (41.1% vs. 84.3%, p < 0.0001).
Conclusions: There is a considerable risk of stent fractures after long segment femoral artery stenting, which is associated with a higher in-stent restenosis and reocclusion rate.