Method for improving the accuracy of quantitative cerebral perfusion imaging

J Magn Reson Imaging. 2005 May;21(5):512-9. doi: 10.1002/jmri.20305.

Abstract

Purpose: To improve the accuracy of dynamic susceptibility contrast (DSC) measurements of cerebral blood flow (CBF) and volume (CBV).

Materials and methods: In eight volunteers, steady-state CBV (CBV(SS)) was measured using TrueFISP readout of inversion recovery (IR) before and after injection of a bolus of contrast. A standard DSC (STD) perfusion measurement was performed by echo-planar imaging (EPI) during passage of the bolus and subsequently used to calculate the CBF (CBF(DSC)) and CBV (CBV(DSC)). The ratio of CBV(SS) to CBV(DSC) was used to calibrate measurements of CBV and CBF on a subject-by-subject basis.

Results: Agreement of values of CBV (1.77 +/- 0.27 mL/100 g in white matter (WM), 3.65 +/- 1.04 mL/100 g in gray matter (GM)), and CBF (23.6 +/- 2.4 mL/(100 g min) in WM, 57.3 +/- 18.2 mL/(100 g min) in GM) with published gold-standard values shows improvement after calibration. An F-test comparison of the coefficients of variation of the CBV and CBF showed a significant reduction, with calibration, of the variability of CBV in WM (P < 0.001) and GM (P < 0.03), and of CBF in WM (P < 0.0001).

Conclusion: The addition of a CBV(SS) measurement to an STD measurement of cerebral perfusion improves the accuracy of CBV and CBF measurements. The method may prove useful for assessing patients suffering from acute stroke.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Blood Flow Velocity
  • Blood Volume Determination
  • Brain / blood supply*
  • Calibration
  • Contrast Media
  • Female
  • Gadolinium DTPA
  • Humans
  • Magnetic Resonance Imaging / methods*
  • Male

Substances

  • Contrast Media
  • Gadolinium DTPA