Precision and accuracy of an ankle-worn accelerometer-based pedometer in step counting and energy expenditure

Prev Med. 2005 Sep-Oct;41(3-4):778-83. doi: 10.1016/j.ypmed.2005.07.006.

Abstract

Background: Walking is a widely used approach to increase physical activity levels in obese patients. In this paper, we investigate the precision and accuracy of an ankle-worn dual-axis accelerometer (Stepwatch) and investigate its potential application as a predictor of energy expenditure.

Methods: Twenty healthy subjects (10 lean, 10 obese) wore spring-levered (Accusplit), piezoelectric (Omron HF-100), and Stepwatch pedometers. Subjects walked on a treadmill at 1, 2, and 3 mph and in a hallway at 1 and 1.85 mph, during which energy expenditure was measured.

Results: The Stepwatch counted 99.7 +/- 0.67% (mean +/- SEM) of the manual counts. In comparison, the Omron pedometer counted 61 +/- 3.3% and the Accusplit counted 26 +/- 2.8% of the manual counts at 1 mph although all pedometers were accurate (> 98% of counts) at 3 mph. In repeated measures, the Stepwatch produced negligible variance (SD = 0.36) over all speed whereas the other pedometers showed a large amount of variance at all speed (SD = 4-13). Stepwatch counts were predictive of walking energy expenditure corrected by weight (r2 > 0.8).

Conclusion: The counts from the Stepwatch were virtually identical to the manual counts from a trained investigator and provided a reliable predictor of walking energy expenditure.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Adult
  • Ankle*
  • Body Mass Index
  • Energy Metabolism / physiology*
  • Equipment Failure Analysis
  • Female
  • Humans
  • Male
  • Middle Aged
  • Minnesota
  • Monitoring, Physiologic / instrumentation*
  • Walking / physiology*