Accuracy and conformity of stereotactically guided interstitial brain tumour therapy using I-125 seeds

Radiother Oncol. 2005 Nov;77(2):202-9. doi: 10.1016/j.radonc.2005.08.006. Epub 2005 Oct 4.

Abstract

Objective: To assess the accuracy of the stereotactic implantation procedure of catheters containing I-125 seeds in brain tumours and investigate the effect of catheter deviations on the dose distribution in patients.

Methods: A randomised sample (n = 37) of all patients treated with I-125 seeds in our department between 6/1994 and 2/2002 was examined. Intraoperative X-ray images were used to measure deviations of implanted I-125 seed catheters from their planned positions and the influence on dose conformity, tumour surface dose and dose burden of surrounding healthy brain tissue was determined.

Results: The mean spatial target point deviation was 2.0 mm (maximum 4.1 mm, SD 0.9 mm) and in 54.1% of the cases, reduction of the planned dose was greater than 5%. Target point deviations less than 1.5 mm have only minor influence on surface dose and conformity. Results indicated that in 10.8% of the cases the realized dose distribution showed a 'slight deviation', according to the guideline criteria for external radiosurgery of the Radiation Therapy Oncology Group. In 89.2% of the patients the applied dose conformed to the target volume.

Conclusions: Stereotactically guided interstitial irradiation with I-125 seeds can be used to treat brain tumours and metastases with high conformity comparable to radiosurgery. The observed deviations of the stereotactically implanted I-125 seed catheters from their planned target points were smaller when compared to frameless procedures. In order to maintain the required spatial accuracy of 1.5 mm in interstitial therapy using I-125 seeds, it appears necessary to optimise stereotactic instruments further.

Publication types

  • Comparative Study
  • Randomized Controlled Trial

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Brain Neoplasms / pathology*
  • Brain Neoplasms / surgery*
  • Child
  • Child, Preschool
  • Dose-Response Relationship, Radiation
  • Female
  • Humans
  • Iodine Radioisotopes / therapeutic use*
  • Magnetic Resonance Imaging
  • Male
  • Maximum Tolerated Dose
  • Middle Aged
  • Neoplasm Staging
  • Radiosurgery / methods*
  • Radiotherapy Planning, Computer-Assisted / methods*
  • Radiotherapy, Computer-Assisted / methods
  • Risk Assessment
  • Sampling Studies
  • Sensitivity and Specificity
  • Stereotaxic Techniques
  • Treatment Outcome

Substances

  • Iodine Radioisotopes