Randomized evaluation of the TriActiv balloon-protection flush and extraction system for the treatment of saphenous vein graft disease

J Am Coll Cardiol. 2005 Nov 1;46(9):1677-83. doi: 10.1016/j.jacc.2005.06.073. Epub 2005 Oct 10.

Abstract

Objectives: The Protection During Saphenous Vein Graft Intervention to Prevent Distal Embolization (PRIDE) study compared outcomes with the TriActiv System (Kensey Nash Corp., Exton, Pennsylvania), a balloon-protection flush and extraction device, with an embolic protection group during treatment of saphenous venous grafts (SVGs).

Background: Treatment of SVGs with embolic protection reduces adverse cardiac events.

Methods: We conducted a prospective trial randomizing 631 patients with coronary ischemia and lesions in SVGs to embolic protection with the TriActiv System or control group (Guardwire System [Medtronic AVE, Santa Rosa, California] or Filterwire EX [Boston Scientific Corp., Maple Grove, Minnesota]).

Results: The incidence of major adverse cardiac events at 30 days was 11.2% for the TriActiv group and 10.1% for the control group (relative risk = 1.1%; 95% confidence interval 0.67 to 1.76; p = 0.65; p = 0.02 for non-inferiority). Safety and efficacy end points were similar between groups except that patients randomized to the TriActiv System had more hemorrhagic complications (10.9% vs. 5.4%; p = 0.01).

Conclusions: The TriActiv System was not inferior to approved embolic protection devices for the treatment of diseased SVGs.

Publication types

  • Randomized Controlled Trial

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Balloon Occlusion / instrumentation*
  • Female
  • Graft Occlusion, Vascular / therapy*
  • Humans
  • Male
  • Middle Aged
  • Postoperative Complications / therapy*
  • Prospective Studies
  • Saphenous Vein / transplantation*