Robustness of late lumen loss in discriminating drug-eluting stents across variable observational and randomized trials

Circulation. 2005 Nov 1;112(18):2833-9. doi: 10.1161/CIRCULATIONAHA105.570093.

Abstract

Background: Binary angiographic and clinical restenosis rates can vary widely between clinical studies, even for the same stent, influenced heavily by case-mix covariates that differ among observational and randomized trials intended to assess a given stent system. We hypothesized that mean in-stent late loss might be a more stable estimator of restenosis propensity across such studies.

Methods and results: In 46 trials of drug-eluting and bare-metal stenting, increasing mean late loss was associated with higher target lesion revascularization (TLR) rates (P<0.001). When the class of bare-metal stents was compared with the class of effective drug-eluting stents, late loss was more discriminating than TLR as measured by the high intraclass correlation coefficient (rho) (late loss, rho=0.71 versus TLR, rho=0.22; 95% CI of difference=0.33, 0.65). When the individual drug-eluting stents and bare-metal stents were compared, late loss was a better discriminator than TLR (0.68 versus 0.19; 95% CI of difference=0.24, 0.60). Greater adjustments of study covariates are needed to stabilize assessments of TLR compared with late loss because of greater influence of reference vessel diameter on TLR than on in-stent late loss. Optimization of late loss with the use of a novel method of standardization according to diabetes prevalence and mean lesion length resulted in minor adjustments in late loss (<0.08 mm for 90% of reported trials) and an ordered array of mean late loss values for the stent systems studied.

Conclusions: Late loss is more reliable than restenosis rates for discriminating restenosis propensity between new drug-eluting stent platforms across studies and might be the optimum end point for evaluating drug-eluting stents in early, nonrandomized studies.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't
  • Review

MeSH terms

  • Blood Vessel Prosthesis Implantation / methods*
  • Clinical Trials as Topic
  • Coronary Restenosis / epidemiology*
  • Coronary Restenosis / prevention & control
  • Drug Implants*
  • Equipment Design
  • Humans
  • Randomized Controlled Trials as Topic
  • Reproducibility of Results
  • Stents*

Substances

  • Drug Implants