Evaluation of a fully automated system (RAISUS) for rapid identification and antimicrobial susceptibility testing of Staphylococci

J Clin Microbiol. 2005 Nov;43(11):5808-10. doi: 10.1128/JCM.43.11.5808-5810.2005.

Abstract

RAISUS is a system for rapid bacterial identification and antimicrobial susceptibility testing. RAISUS and VITEK showed 97.8% and 75.9% agreement in identification of 45 Staphylococcus aureus strains and 58 coagulase-negative staphylococci (CoNS), respectively, and RAISUS and CLSI (formerly NCCLS) methods showed 87.2% and 87.9% agreement in the MICs for S. aureus and CoNS, respectively. RAISUS provided these data within 3.75 h, suggesting its utility for clinical bacteriological laboratories.