High sensitivity determination of valproic acid in mouse plasma using semi-automated sample preparation and liquid chromatography with tandem mass spectrometric detection

Rapid Commun Mass Spectrom. 2005;19(24):3713-8. doi: 10.1002/rcm.2249.

Abstract

A high-throughput liquid chromatography/electrospray ionization tandem mass spectrometry (LC/ESI-MS/MS) assay using automated sample preparation has been developed for the determination of valproic acid (VPA) in mouse plasma. A liquid-handling system was programmed to prepare calibration standard solutions in plasma, as well as quality controls and clinical samples. Plasma protein precipitation was performed on a 96-well plate, and the collected supernatant was directly injected into a reversed-phase LC/ESI-MS/MS system in the negative ionization mode. The calibration curve for VPA was linear over a dynamic range of 0.15-100 microg/mL. The limit of detection was 75 ng/mL and the lower limit of quantitation was 150 ng/mL. Intra- and inter-day validation assays of the semi-automated plasma analysis showed satisfactory accuracy and precision.

MeSH terms

  • Animals
  • Automation
  • Chromatography, Liquid / instrumentation
  • Chromatography, Liquid / methods*
  • Mass Spectrometry / instrumentation
  • Mass Spectrometry / methods*
  • Mice
  • Sensitivity and Specificity
  • Valproic Acid / blood*

Substances

  • Valproic Acid