Accuracy and reliability of remote retinopathy of prematurity diagnosis

Arch Ophthalmol. 2006 Mar;124(3):322-7. doi: 10.1001/archopht.124.3.322.

Abstract

Objective: To determine the accuracy and reliability of retinopathy of prematurity (ROP) diagnosis using remote review of digital images by 3 masked ophthalmologist readers.

Methods: An atlas was compiled of 410 retinal photographs from 163 eyes of 64 low-birth-weight infants taken using a wide-angle digital fundus camera. All the images were independently reviewed by 3 readers, and the diagnosis in each eye was classified into 1 of 4 ordinal categories: no ROP, mild ROP, type 2 prethreshold ROP, or ROP requiring treatment. Findings were compared with a reference standard of dilated indirect ophthalmoscopy with scleral depression performed by an experienced pediatric ophthalmologist.

Results: Sensitivities/specificities of the diagnosis of any ROP were 0.845/0.910 for the first reader, 0.816/0.955 for the second reader, and 0.864/0.493 for the third reader. Sensitivities/specificities of the diagnosis of ROP requiring treatment were 0.850/0.960 for the first reader, 0.850/0.973 for the second reader, and 0.900/0.953 for the third reader. When ROP was classified into ordinal categories, the overall weighted kappa for interreader reliability was 0.743. Intrareader reliability for detection of low-risk prethreshold ROP or worse was 100% for all readers.

Conclusion: The accuracy, interreader reliability, and intrareader reliability of remote diagnosis of clinically relevant ROP based on digital imaging are substantial.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Birth Weight
  • Diagnostic Techniques, Ophthalmological*
  • Gestational Age
  • Humans
  • Infant, Newborn
  • Infant, Very Low Birth Weight
  • Observer Variation
  • Ophthalmoscopy
  • Photography / methods*
  • Reference Standards
  • Remote Consultation*
  • Reproducibility of Results
  • Retinopathy of Prematurity / classification
  • Retinopathy of Prematurity / diagnosis*
  • Sensitivity and Specificity