Multislice dark-blood carotid artery wall imaging: a 1.5 T and 3.0 T comparison

J Magn Reson Imaging. 2006 May;23(5):699-705. doi: 10.1002/jmri.20563.

Abstract

Purpose: To compare two multislice turbo spin-echo (TSE) carotid artery wall imaging techniques at 1.5 T and 3.0 T, and to investigate the feasibility of higher spatial resolution carotid artery wall imaging at 3.0 T.

Materials and methods: Multislice proton density-weighted (PDW), T2-weighted (T2W), and T1-weighted (T1W) inflow/outflow saturation band (IOSB) and rapid extended coverage double inversion-recovery (REX-DIR) TSE carotid artery wall imaging was performed on six healthy volunteers at 1.5 T and 3.0 T using time-, coverage-, and spatial resolution-matched (0.47 x 0.47 x 3 mm3) imaging protocols. To investigate whether improved signal-to-noise ratio (SNR) at 3.0 T could allow for improved spatial resolution, higher spatial resolution imaging (0.31 x 0.31 x 3 mm3) was performed at 3.0 T. Carotid artery wall SNR, carotid lumen SNR, and wall-lumen contrast-to-noise ratio (CNR) were measured.

Results: Signal gain at 3.0 T relative to 1.5 T was observed for carotid artery wall SNR (223%) and wall-lumen CNR (255%) in all acquisitions (P < 0.025). IOSB and REX-DIR images were found to have different levels of SNR and CNR (P < 0.05) with IOSB values observed to be larger. Normalized to a common imaging time, the higher spatial resolution imaging at 3.0 T and the lower spatial resolution imaging at 1.5 T provided similar levels of wall-lumen CNR (P = NS).

Conclusion: Multislice carotid wall imaging at 3.0 T with IOSB and REX-DIR benefits from improved SNR and CNR relative to 1.5 T, and allows for higher spatial resolution carotid artery wall imaging.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Blood
  • Carotid Arteries / anatomy & histology*
  • Feasibility Studies
  • Female
  • Humans
  • Image Enhancement / instrumentation
  • Image Enhancement / methods*
  • Image Processing, Computer-Assisted / methods
  • Magnetic Resonance Imaging / instrumentation
  • Magnetic Resonance Imaging / methods*
  • Magnetics*
  • Male
  • Reference Values