Half-life measurement of 55Fe

Appl Radiat Isot. 2006 Oct-Nov;64(10-11):1412-6. doi: 10.1016/j.apradiso.2006.02.052. Epub 2006 Mar 30.

Abstract

The half-life of 55Fe was determined experimentally and compared with literature values. A solid 55Fe source was prepared by electrodeposition and repeatedly measured in an X-ray detector at a constant, low solid angle. In total, more than six hundred measurements were performed over a period of ten months, corresponding to one third of the half-life. The experimental method and the corresponding uncertainty budget are presented. The result of 1005.2+/-1.4 days is slightly higher than the current recommended values.

Publication types

  • Evaluation Study

MeSH terms

  • Algorithms
  • Computer Simulation
  • Guidelines as Topic
  • Half-Life
  • Iron Radioisotopes / analysis*
  • Iron Radioisotopes / chemistry*
  • Iron Radioisotopes / standards
  • Models, Chemical*
  • Radiation Dosage
  • Radiometry / instrumentation
  • Radiometry / methods*
  • Reference Standards
  • Reference Values

Substances

  • Iron Radioisotopes