Objective: To evaluate the clinical performance of the Pediatric Vision Screener (PVS) in children in a pediatric ophthalmology office setting.
Methods: Seventy-seven subjects between 2 and 18 years of age received gold-standard orthoptic examinations and were classified as at risk for amblyopia if strabismus or anisometropia (>1.50 diopters) was present. Strabismus was subclassified as variable or constant. The subjects were then tested with the PVS, which produced a pass or refer recommendation based on a binocularity score. The PVS also produced a yield score to indicate the subject's interest in the target. Sensitivity and specificity for amblyopia risk detection were calculated.
Results: Binocularity as determined by the PVS was greater than 65% for all controls and less than 20% for all subjects with constant strabismus. Binocularity ranged from 0% to 52% in subjects with variable strabismus. All subjects with anisometropia and no strabismus had binocularity scores less than 10%.
Conclusions: The PVS identified strabismus, when present, in all subjects and identified 3 subjects with anisometropia as well. The PVS shows potential to address a lack of screening instrumentation appropriate for use with preschool-aged children.