A method is presented for the calculation of perfusion parameters in dynamic contrast enhanced MRI. This method requires identification of enhancement curves for both tumor tissue and plasma. Inter-operator variability in the derived rate constant between plasma and extra-cellular extra-vascular space is assessed in both canine and human subjects using semi-automated tumor margin identification with both manual and automated arterial input function (AIF) identification. Experimental results show a median coefficient of variability (CV) for parameter measurement with manual AIF identification of 21.5% in canines and 11% in humans, with a median CV for parameter measurement with automated AIF identification of 6.7% in canines and 6% in humans.