High frequency broadband PZT thick film ultrasonic transducers for medical imaging applications

Ultrasonics. 2006 Dec 22:44 Suppl 1:e711-5. doi: 10.1016/j.ultras.2006.05.130. Epub 2006 Jun 6.

Abstract

A modified sol-gel method is used to prepare PZT thick film on Pt-coated silicon substrate. A new method of vacuum filling sol-gel precursor solution is introduced to improve film quality. The effects of the filling on PZT thick film structure and ferroelectric properties are discussed. The fabrication of a high frequency transducer with the PZT film as the actuating layer is described. The performance of the transducer is measured and results show that the transducer backed by E-Solder without a matching layer has a center frequency of 103 MHz and a bandwidth of 70%. Beam profile measurements show that the transducer has an axial resolution of 9.2 microm and a lateral resolution of 33 microm.

Publication types

  • Evaluation Study
  • Research Support, N.I.H., Extramural

MeSH terms

  • Electromagnetic Phenomena / instrumentation*
  • Electromagnetic Phenomena / methods
  • Equipment Design
  • Equipment Failure Analysis
  • Membranes, Artificial*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Transducers*
  • Ultrasonography / instrumentation*
  • Ultrasonography / methods

Substances

  • Membranes, Artificial