Standardization of 222Rn by LSC and comparison with alpha- and gamma-spectrometry

Appl Radiat Isot. 2006 Oct-Nov;64(10-11):1465-70. doi: 10.1016/j.apradiso.2006.02.068. Epub 2006 Jul 28.

Abstract

Liquid scintillation counting (LSC) was used for the measurement of 222Rn in equilibrium with its daughters, with detection efficiency close to 5. The appropriate corrections were considered, including one related to the probability that the 165-micros half-life 214Po decays during the dead time of the counter initiated by the disintegration of his parent nuclide, 214Bi. The dead-time determination of a commercial LS counter is also presented using a 222Rn standard source. The LSC 222Rn sources were prepared by transfer of 222Rn produced by a solid 226Ra source into LSC cocktail frozen at 77K, flame-sealed afterwards. They were measured using the LNHB triple coincidence counter with adjustable extending-type dead-time unit, between 8 and 100 micros; two different procedures were used to calculate an effective dead time and then to deduce the counting rate extrapolated to zero dead-time value. The LSC results were compared with those obtained by cryogenic alpha-particle spectrometry (LNHB system) and by gamma-ray spectrometry for the same radon source in the LSC vial; the geometry transfer coefficient was calculated using the ETNA software. Measurement results and uncertainties are discussed.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms*
  • Alpha Particles*
  • Computer Simulation
  • France
  • Models, Chemical*
  • Radiation Dosage
  • Radon / analysis*
  • Radon / standards*
  • Romania
  • Scintillation Counting / methods*
  • Scintillation Counting / standards*
  • Spectrometry, Gamma / methods*
  • Spectrometry, Gamma / standards

Substances

  • Radon