Primary stability of a conical implant and a hybrid, cylindric screw-type implant in vitro

Int J Oral Maxillofac Implants. 2006 Jul-Aug;21(4):560-6.

Abstract

Purpose: The differences with respect to primary stability between 2 Camlog implants, a conical implant, and a hybrid cylindric screw-type implant, were investigated in vitro. The effect of underdimensioned implant bed preparation was also studied for both implant designs.

Materials and methods: In an in vitro model the stability of different implants in fresh porcine iliac bone blocks was measured using torque moment values, the Periotest, resonance frequency analysis, and push-out testing.

Results: The conical implant showed significantly higher primary stability than the cylindric hybrid implant using the insertion torque, Periotest, and push-out tests. For both types of implants, the torque moment values following under-dimensioned preparation were significantly better than those obtained following the standard drilling protocol (Conical: 25.00 vs 11.00 Ncm; Cylindrical: 11.75 vs. 5.75 Ncm). For the cylindric implant, significantly better results following under-dimensioned implant bed preparation were observed only with the insertion torque and the pushout testing values. The mean ISQ values for all groups were between 55 and 57; no statistical differences with respect to ISQ could be found.

Conclusion: In this in vitro model conical implants showed higher primary stability than cylindric implants. The procedure of under-dimensioned drilling seemed to increase primary stability for both types of implants; however, the effect was only observable using insertion torque. RFA and Periotest, the noninvasive, clinical methods tested, did not clearly demonstrate this difference.

MeSH terms

  • Animals
  • Dental Implantation, Endosseous / methods*
  • Dental Implants*
  • Dental Prosthesis Design*
  • Dental Prosthesis Retention*
  • Swine
  • Torque