Hemodynamic energy generated by a combined centrifugal pump with an intra-aortic balloon pump

ASAIO J. 2006 Sep-Oct;52(5):592-4. doi: 10.1097/01.mat.0000237692.12805.01.

Abstract

We examined the pulsatility generated by an intra-aortic balloon pump/centrifugal pump (IABP/CP) combination in terms of energy equivalent pressure (EEP) and surplus hemodynamic energy (SHE). In five cardiac-arrested pigs, the outflow cannula of the CP was inserted into the ascending aorta, the inflow cannula in the right atrium. A 30-ml IABP was subsequently placed in the descending aorta. Extracorporeal circulation was maintained for 30 minutes using a pump flow of 75 ml/kg per minute by CP alone or by IABP/CP with pressure and flow measured in the right internal carotid artery. The IABP/CP combination converted the flow to pulsatile and increased pulse pressure significantly from 9.1 +/- 1.3 mm Hg to 54.9 +/- 6.1 mm Hg (p = 0.012). It also significantly increased the percent change from mean arterial pressure to EEP from 0.2 +/- 0.3% to 23.3 +/- 6.1% (p = 0.012) and SHE from 133.2 +/- 234.5 erg/cm to 20,219.8 +/- 5842.7 erg/cm3 (p = 0.012). However, no statistical difference was observed between CP and IABP/CP in terms of mean carotid artery pressure (p = NS). In a cardiac-arrested animal model, pulsatility generated by a IABP/CP combination may be effective in terms of energy equivalent pressure and surplus hemodynamic energy.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Cardiac Catheterization / methods
  • Cardiovascular Physiological Phenomena*
  • Heart-Assist Devices*
  • Intra-Aortic Balloon Pumping / instrumentation*
  • Swine