Reduction of atrial tachyarrhythmia episodes during the overdrive pacing period using the post-mode switch overdrive pacing (PMOP) algorithm

Heart Rhythm. 2006 Oct;3(10):1164-71. doi: 10.1016/j.hrthm.2006.06.007. Epub 2006 Jun 15.

Abstract

Background: Early recurrences within 10 minutes after termination of an atrial tachyarrhythmia, such as atrial tachycardia (AT), atrial flutter, or atrial fibrillation (AF) episodes, account for a large part of paroxysmal AT/AF episodes. It is unclear whether these arrhythmias can be suppressed by overdrive pacing.

Objective: We set out to prove that overdrive pacing is effective in preventing early recurrences of AT/AF.

Methods: This AT500 (DDDRP device, Bakken Research Center Medtronic, Maastricht, The Netherlands) post-mode switch overdrive pacing (PMOP) study is a randomized controlled trial designed to test the efficacy of overdrive pacing on the suppression of early recurring AT/AF episodes. With the PMOP feature, overdrive pacing is activated 12 ventricular beats after device-confirmed termination of an AT/AF episode with a programmed duration and rate. If at least four episodes of 1 minute in duration occurred within the run-in period of 1 month, patients were randomized to one of the three settings (PMOP off and PMOP 10 minutes at 90 bpm or 120 bpm) for 2 months. At 2 and 4 months, patients were crossed over to another arm. At 6 months, all patients were programmed with PMOP on at 90 bpm for 30 minutes.

Results: We enrolled 122 and randomized 50 patients. Sixty percent of all stored AT/AF episodes occurred within 10 minutes after a previous episode; 31% occurred after device-determined termination and before the device reached the overdrive rate (17-27 ventricular beats after termination), and 29% of the episodes occurred while the device was pacing at the programmed overdrive rate. Thirty-seven percent of the average percentage of episodes during the overdrive period was prevented by the randomized settings of PMOP 90 bpm/10 min (P = .01, paired t-test, n = 39) and 120 bpm/10 min (P = .03, n = 35). In addition, for the nonrandomized setting of PMOP 90 bpm/30 min, the average number of episodes during the overdrive period was significantly smaller than the percentage of episodes occurring during the corresponding off period of 30 minutes (P = .05, n = 33). No significant differences in burden and episodes were found between the PMOP settings and the corresponding off periods.

Conclusions: This is the first randomized controlled clinical trial to prove that overdrive pacing is effective in preventing early recurrences of AT/AF. However, shortcomings of the PMOP algorithm, such as late activation, need further improvement.

Publication types

  • Randomized Controlled Trial

MeSH terms

  • Aged
  • Aged, 80 and over
  • Algorithms*
  • Cardiac Pacing, Artificial / methods*
  • Follow-Up Studies
  • Heart Rate / physiology
  • Humans
  • Middle Aged
  • Prospective Studies
  • Quality of Life
  • Secondary Prevention
  • Single-Blind Method
  • Tachycardia, Ectopic Atrial / physiopathology
  • Tachycardia, Ectopic Atrial / psychology
  • Tachycardia, Ectopic Atrial / therapy*
  • Treatment Outcome